Brand Name :GUOGUANG
Model Number :ZDF-5227
Place of Origin :Sichuan, China (Mainland)
MOQ :Negotiation
Price :Negotiation
Payment Terms :T/T
Supply Ability :200 Pieces per Month
Delivery Time :10 days(less than 50pcs)
Packaging Details :Export Packing: Carton box
Pressure Range :1.0×105—1.0×10-5 (Pa)
Testing precision :±10%
Equipped gauge :resistance gaugeZJ-52T/metal , ionization gauge ZJ-27/glass
Control precision :±1%
Control time :< 1S
Output contend :3A/220VAC non-inductive load
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

Thermocouple Vacuum Controller ZDF-5227

Thermocouple Vacuum Controller:

ZDF-5227 compound vacuum gauge, which is composed by resistance meter and ionization meter, can test signs of the vacuum gauge with two circuits and has two circuits control contact output. In the testing range, it can be point control and area control freely, and it is suitable to be used in testing and controlling the vacuum of wide measuring range.

It adopts the most advanced microprocessor as its main machine, improving its reliability by decreasing external components. Its hardware employs the modular structure, EMI restraint and WATCHDOG circuit; while the software employs anti-interference technique, such as redundancy, trap, digital filtering and so on to improve its anti-interference ability and performance, which ensures the high reliability of the apparatus and its normal use in bad industrial environment.

Parameter

1

Testing circuit quantity

Two circuits

2

Testing range

(The equipped vacuum gauge)

1.0×105—1.0×10-5 (Pa)

(resistance gaugeZJ-52T/metal , ionization gauge

ZJ-27/glass)

3

Ionization parameters

Voltage of Positive pole:225V

Emission current:1mA/100uA

4

Testing precision

±10%

5

Zero drift

±2%

6

Controlling range

1.0×105—1.0×10-5 (Pa)

7

Controlling method

Two circuits point control or area control

8

Controlling precision

±1%

9

Controlling time

< 1S

10

Output contend

3A/220VAC non-inductive load

11

Analog output

0-5V/4-20m

12

Power supply

AC 220V/50Hz ±10%

13

Environmental temperature

-10℃--45℃

14

Environmental Humidity

< 85%(25℃)

15

Rated consumed power

35W

16

Net weight of the apparatus

3.5Kg around